Krios Status

Legend

Refreshing

 
Refresh in:
Back to systems overview

Overview

ParameterActual valueExpected value
System type Local FEI Titan Cubed
Microscope connection Ready
Alarms raised 0 0
Alarms pending 0 0
Last update 2 min ago < 2.5 min ago

Vacuum

ParameterActual valueExpected value
Vacuum state Ready
Gun pressure 1 Log < 2 Log
Column pressure 11 Log < 15 Log
Camera pressure 30 Log < 35 Log
Column valves Open

Gun

ParameterActual valueExpected value
High tension state On
High tension value 300 kV 50 - 300 kV

Misc Microscope

ParameterActual valueExpected value
Stage state Moving
Hand panels connection Connected

Chillers

ParameterActual valueExpected value
Chiller state OK
Chiller input temperature 7.9 °C 7.5 - 14.0 °C
Chiller output temperature 18.5 °C 16.7 - 20.0 °C

PC and Network

ParameterActual valueExpected value
Windows user IBC-37
Free space on C:\ 106 GB > 50 GB
Connectivity to 198.211.143.66 OK

Environment

ParameterActual valueExpected value
Flood alarm Krios OK
Room temperature Krios 22.6 °C 16.0 - 26.0 °C
Humidity Krios 18% rH 10 - 70% rH
Flood alarm Chiller OK
Room temperature Chiller 10.4 °C 7.0 - 23.0 °C
Humidity Chiller 40% rH

Nexperion Sentinel

ParameterActual valueExpected value
Mail message queue 0 < 5
SMS message queue 0 < 5
Uptime 21 days
Software version 23.08.0 23.08.0
Remote address 192.168.6.77
Generated by Nexperion Sentinel Status 23.01.0 in 60.34 milliseconds