TEM Status

Legend

Refreshing

 
Refresh in:

Overview

ParameterActual valueExpected value
System type Generic 200 kV TEM
Microscope connection Ready
Alarms raised 3 0
Alarms pending 3 0
Last update 2 min ago < 2.5 min ago

Vacuum

ParameterActual valueExpected value
Vacuum state Ready
Gun pressure 1.808E-07 Pa < 8.009E-06 Pa
Column pressure 7.821E-05 Pa < 3.000E-05 Pa
Camera pressure 3.770E-05 Pa < 2.500E-04 Pa
Column valves Closed

Gun

ParameterActual valueExpected value
High tension state On
High tension value 80 kV 80 - 200 kV

Misc Microscope

ParameterActual valueExpected value
Stage state Idle
Hand panels connection Connected

Chillers

ParameterActual valueExpected value
Chiller state OK
Chiller input temperature 8.9 °C 4.0 - 11.5 °C
Chiller output temperature 16.3 °C 16.0 - 18.0 °C

Media

ParameterActual valueExpected value
Pressure Compressed air 6.9 bar 5.5 - 8.0 bar
Pressure N2 0.2 bar 0.1 - 0.5 bar

PC and Network

ParameterActual valueExpected value
Windows user
Free space on C:\ 226 GB > 150 GB
Free space on D:\ 962 GB > 50 GB
Free space on \\ImageServer 48 GB > 100 GB
Connectivity to ImageServer OK

Environment

ParameterActual valueExpected value
Flood alarm Microscope OK OK
Room temperature Microscope 22.3 °C 21.0 - 23.0 °C
Humidity Microscope 46% rH 0 - 30% rH
Flood alarm Technics OK OK
Room temperature Technics 23.4 °C 20.0 - 25.5 °C
Humidity Technics 42% rH 0 - 60% rH

Nexperion Sentinel

ParameterActual valueExpected value
Mail message queue 0 < 5
SMS message queue 0 < 5
Uptime 3 weeks
Software version 23.08.0 23.08.0
Remote address 503.hosttech.eu
Generated by Nexperion Sentinel Status 23.01.0 in 17.14 milliseconds
© 2014 – 2025 Nexperion e.U. – Solutions for Electron Microscopy
For feedback, contact nexperion(at)nexperion.net